HAEMATOKRIT 210, bench-top centrifuge without rotorSpecial-purpose CentrifugeAt 16,060 g sedimentation is completed in just a little over 6 minutes. At any rate, a run time of 10 minutes is sufficient. The rotor comes to a braked standstill. Each capillary is secured within its own chamber and supported by a holding tray. This holding tray serves to contain glass shards and any leakage should a tube break. It is easy to clean and disposable the case of glass breakage.
Capillaries, to be used for quantitative buffy coat analysis are centrifuged in the 20-place rotor prior to further analysis.
User friendless:
Quick-entry keypad: easy-to-read, digital display, display of current parameter valuesEasy entry of the parameters: RPM in increments of 100, running time in minutes, max. 99 minSafety:
Lid locking and holdingEmergency lid lock releaseMotor overheating protectionImbalance switch-offStandstill indicationTechnology:
Consumption 300 VA(27302104) 220 VA(27302104-01)Radio interference EN 55011,ISM class B (27302104) FCC class B (27302104-01)Immunity EN 50082-1 (27302104)Max. capacity 24/20 capillariesMax. RPM (speed) 13,000 /minMax. RCF 16,060Dimensions (H x W x D) in mm 247 x 275 x 330Weight approx. 10 kg